3-19
VIBXPER
T II 05.2012
Left:
Changing the measurement task
Right:
Task Manager
Setups for
measurement, sensor, evaluation and
RPM measurement
* Preconfigured measurement tasks/set-
ups are marked with a lock and cannot
be edited (except for transducer
& channel).
**Initially only the 'available transduc-
ers' appear in the transducer setup
(see P. 2-19). To display all transducers,
click on the bottom list entry '...more
transducers'
Open the 'Task Manager '
M = in Multimode only
Options
Use this option when you do not have an RPM sensor on the route and
need to perform measurements that require the RPM (e.g. roller bear-
ing condition, order-based spectra). If the measurement task already
has information on the RPM, this option is ignored.
Changing the measurement task (M)
• Highlight the measurement type in the selection screen.
• Press the F key. A list appears with all the measurement tasks set
up for this measurement type.
• Click on the required measurement task (see below).
Setting up a new measurement task (M)
VIBXPERT has a large selection of frequently used measurement
tasks* that can be extended by user-defined measurement tasks if
necessary. To configure a new measurement task, proceed as follows:
• Highlight the measurement type in the selection screen.
• Press the MENU key and click on 'Task Manager'. The Task Man-
ager appears and can be used to set up the new measurement task
(see P. 3-2).
• Click on the 'Measurement task' pulldown menu.
• Press the MENU key and click on 'New'. The text editor appears.
• Enter a name for the new measurement task.
• Change the measurement channel if necessary.
• Select the setups for measurement (see below), sensor**, evalua-
tion or RPM in the 'Setup' field of the pulldown menu. If the match-
ing entry cannot be found, a new setup must be created (see next
section).
• Then press the MENU key and click on 'OK'.