Getting started - Basic settings
Left:
Basic settings for the route/tem-
plate operating mode
Right:
Options for saving
Standard vibration sensor
Here you define the sensor to be used for all vibration measurements
on a route. The settings in OMNITREND are ignored. The 'Triax sen-
sor' option is disabled.
Standard sensor is active when, in the task selection screen
- Channel A / Channel B is marked red,
- Sensor icon at the top edge of the screen lights up orange.
Use the standard sensor for measuring or deactivate this option.
Otherwise, incorrect measurements may be the result.
Save settings
• Click on 'Saving' to set up the following options:
AutoSave
This shortens the duration of the entire measurement.
-
e
naBled
: The selected time (0 ... 10s) corresponds to the waiting
time until the result is automatically saved.
-
d
isaBled
: The result must be manually saved.
Stop AutoSave on...
-
a
larm
, w
arning
, p
rewarning
: When the limit value is exceeded, the
AutoSave mode is deactivated.
-
n
ever
: The result is always stored automatically.
To accept the settings in the 'Route Setup' and to close the menu, press
MENU and click on 'OK'.