5-26
VIBXPER
T II 05.2012
Left:
Reference measurement
is required when measuring on a run-
ning machine
Right:
Dataset selection
Single measurements /
Averaged final result
Measurement tasks
Tips for setting the measurement parameters
Set the 'Trigger Start" time to a negative value in order to record the
signal before the excitation.
If the measurement starts before the excitation, increase the trig-
ger threshold in the measurement setup by approximately half (P.
3-22).
If the response signal overloads, increase the 'Measurement Range'
parameter in the measurement setup accordingly (P. 3-22).
Note
Measurement with the machine running
• Enable 'Negative averaging' in the measurement setup.
• Perform the individual impact tests.
• Click on 'Start' in the window that opens to measure the reference
spectrum - without producing a machine excitation!
• Click on 'Stop' when a sufficient number of reference spectra have
been recorded and averaged to form a result (see below).
Displaying the results
• Activate the upper pane with the F key (Spectrum, P. 4-8).
• Press the MENU key and click on 'Choose Dataset'.
The results of the impact test, the average final result and the refer-
ence measurement (negative averaging) are listed here and can be
displayed individually by clicking (see below).
• To close the list, press on the ESC key.
Repeating individual measurements
• Activate the upper pane using the F key.
• Press the MENU key and click on 'Repeat'.
• Click on the respective measurement in the list.