3-22
VIBXPER
T II 05.2012
Options
Measurement setup for
- Runup/coastdown analysis
(left)
- Orbit
(right)
*'Fixed range', e.g. for a runup / coast-
down curve or impact test.
Tip: Carry out a test measurement
in order to determine the maximum
amplitude.
o
rder
(
with
p
hase
, o
rBit
, c
oastdown
measurement
)
Order filter for the harmonic vibration components of the signal.
For example, only the signal components of the 1st harmonic are
filtered out and displayed by the first order filter. A half order is
used for detecting oil whirls in machines with journal bearings.
p
ulse
per
rev
. (
with
rpm
measurement
):
Number of measurement marks
on the shaft or coupling.
m
easurement
range
(
all
viBration
measurements
):
The measurement
range for the analog channel (A/B) can be automatically matched
to the input signal ('Auto') or set to fixed*. When set to ‘[Value] /
Auto-Up’, the measurement range is automatically increased when
the upper limit is exceeded.
l
ow
pass
filter
(p
hase
, a
mplitude
c
oast
-
down
- p
hase
):
To suppress high
frequency noise signals, a low pass filter (1 kHz) can be activated
(Yes). In a coastdown measurement the low pass filters (1 kHz /
10 kHz) are set automatically as a function of the start and stop
speeds. With 'No', the signal is processed without a low pass filter.
w
indow
/ n
umBer
of
lines
(i
mpact
test
):
The standard window function
for the impact test with an impulse hammer is the rectangular win-
dow; all other parameters are described in section A2.
n
egative
a
veraging
(i
mpact
t
est
):
Enables an impact test while the ma-
chine is running (1 channel and 2 channels with impulse hammer).
The signals from the running machine are filtered out.
frf
calculation
(2-
ch
impact
test
):
Calculation procedure for the trans-
fer function - (Frequency Response Function = FRF).
t
rigger
type
/
level
/
start
(
s
hock
p
ulse
, v
iBration
overall
v
alue
, u
ser
ac, i
mpact
t
est
)
Level: The measurement starts as soon as the signal has exceeded
the selected trigger level (in % of the max. meas. range). Signal
recording can be started before or after the trigger event via the
Trigger start time.