Startup . 53
You should follow the on-screen instructions to remove the SPME.
6.
Immediately depress the plunger on the SPME syringe to retract the fiber into the Custodion
needle and remove the syringe.
CAUTION
ATTENTION
Ensure the SPME fiber is completely w ithdraw n inside the
syringe needle, before removing the Custodion from the injector
port. Failure to do so may result in the SPME fiber being
damaged and may also cause a helium leakage.
Assurez-vous que la fibre SPME est complètement retirée à
l'intérieur de l'aiguille de la seringue, avant de retirer la
Custodion du port de l'injecteur. Le non-respect de cette
consigne peut entraîner l'endommagement de la fibre SPME et
provoquer également une fuite d'hélium.
The following cutaway diagram shows the Custodion syringe inserted into the injector port
with fiber retracted.
(1) SMPE Syringe
(2) Interface between Custodion syringe and
injector
(3) Injector (hidden under face plate)
(4) Inlet of GC column
A screen similar to the following displays a chromatogram while running a sample.
Содержание TORION T-9
Страница 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Страница 4: ...iv Torion T 9 User s Guide...
Страница 10: ...6 Torion T 9 User s Guide...
Страница 11: ...1 Introduction...
Страница 15: ...2 Safety Practices...
Страница 24: ...20 Torion T 9 User s Guide...
Страница 31: ...3 System Overview...
Страница 40: ...36 Torion T 9 User s Guide...
Страница 41: ...4 Basic Custodion Operations...
Страница 49: ...5 Basic Operation...
Страница 116: ...112 Torion T 9 User s Guide...
Страница 117: ...6 Software...
Страница 257: ...7 Advanced Operations...
Страница 260: ...256 Torion T 9 User s Guide...
Страница 261: ...8 Maintenance...
Страница 268: ...264 Torion T 9 User s Guide...
Страница 269: ...9 Troubleshooting...