160 . Torion T-9 User’s Guide
•
None
indicates that the method will not attempt to calculate a quantitation value
for the compound.
•
Target Compound
indicates that the method is meant to calculate a concentration
for the compound.
•
Internal Standard
indicates that the compound is an internal standard to be used
to calculate the concentration of the target compounds.
e)
Reference Compound
is used to select an internal standard compound as the
designated standard for the target compound. This allows for selection from any of the
designated internal standard compounds. When a compound has its type set to
Internal Standard
it is automatically added to the reference compound list. There is
no limit to the number of compounds that can be designated as internal standards.
To select a
Reference Compound
click on the arrow on the
Reference Compound
box in the row of the compound that is to be the target compound and then click on the
name of the internal standard peak. The following screen shows a drop down where
Methyl Salicylate and Toluene-d8 have been setup as internal standards. By clicking on
one of the compound names that compound will be designated as the internal standard
peak for the Chlorobenzene. They both appear in the dropdown menu under reference
compound.
f)
The
Hazard Level
is a self-configured value. The information displayed in the
Hazard
Level
column can be configured by selecting the
Edit
menu from the menu bar at the
top of the screen and selecting the
Hazard
Level
option. Information about entering
information into a
Hazard
Level
table is given in the menu section of this chapter
under
Edit/Hazard Level
.
To select a
Hazard Level
Click the drop-down arrow and select the desired Hazard
Level from the list.
Since this data is custom created it can be setup to display any information desired. For
example it can be used simply as a “present” or “not present” option for applications
where the chemicals being analyzed are not considered hazardous.
Содержание TORION T-9
Страница 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Страница 4: ...iv Torion T 9 User s Guide...
Страница 10: ...6 Torion T 9 User s Guide...
Страница 11: ...1 Introduction...
Страница 15: ...2 Safety Practices...
Страница 24: ...20 Torion T 9 User s Guide...
Страница 31: ...3 System Overview...
Страница 40: ...36 Torion T 9 User s Guide...
Страница 41: ...4 Basic Custodion Operations...
Страница 49: ...5 Basic Operation...
Страница 116: ...112 Torion T 9 User s Guide...
Страница 117: ...6 Software...
Страница 257: ...7 Advanced Operations...
Страница 260: ...256 Torion T 9 User s Guide...
Страница 261: ...8 Maintenance...
Страница 268: ...264 Torion T 9 User s Guide...
Страница 269: ...9 Troubleshooting...