Method Menu . 139
3.
Type a
Method name
and then click
OK
.
By clicking the
OK
button a new method is created in the database.
After entering a method name and clicking
OK
two verification dialogs display. The first
message that displays is: “The Performance Validation (calibration) data in this method is
not valid. Do you want to load default settings?”
4.
If the response is
Yes
a second information dialog will display with the message: “The
default values have been loaded. For Performance Validation to work correctly it may be
necessary to adjust the mass and/or the retention time calibration windows.” Click OK to
dismiss the window.
5.
If the response is
No
a different dialog box will display with the message: “It will be
necessary to manually configure the Performance Validation settings before using this
method to calibrate the instrument.” Click OK to dismiss the window.
After creating the new method the Method settings forms in the navigation pane will
become enabled. Three specific forms,
Gas Chromatograph
,
Target Library Editor
, and
Performance Validation
, are used to adjust settings to the method.
Содержание TORION T-9
Страница 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Страница 4: ...iv Torion T 9 User s Guide...
Страница 10: ...6 Torion T 9 User s Guide...
Страница 11: ...1 Introduction...
Страница 15: ...2 Safety Practices...
Страница 24: ...20 Torion T 9 User s Guide...
Страница 31: ...3 System Overview...
Страница 40: ...36 Torion T 9 User s Guide...
Страница 41: ...4 Basic Custodion Operations...
Страница 49: ...5 Basic Operation...
Страница 116: ...112 Torion T 9 User s Guide...
Страница 117: ...6 Software...
Страница 257: ...7 Advanced Operations...
Страница 260: ...256 Torion T 9 User s Guide...
Страница 261: ...8 Maintenance...
Страница 268: ...264 Torion T 9 User s Guide...
Страница 269: ...9 Troubleshooting...