86 . Torion T-9 User’s Guide
5.
Enter an appropriate
Multiplier
if the actual sampling conditions are different from what is
specified in the Method.
For example, if the Method specifies sample collection of 1.0 L of air, but you were only able
to collect 0.5 L, you would enter a multiplier of 2.00 (1.0 L / 0.5 L).
6.
Press the
Start Run
button and follow the on-screen prompts.
The next screen shows how the system displays the concentration results for a Quantitative
Method. The concentration units displayed in the
CONC
column are defined in the method
using the Chromion software.