208 . Torion T-9 User’s Guide
Library Search
The
Library Search
option is only available by right clicking on a spectra chart. The
Library
Search Spectra
Graph
shows the results of a library search and therefore is not used to initiate a
library search.
The library to be searched and the RI window size are set in the
Deconvolution Parameters
tab
in Target Library Editor. Refer to the section of this chapter that describes the deconvolution settings
tab for instructions on how to select a library and how to set up the value for the windows.
Selecting the
Library Search…
option opens the following dialog box.
Calculated Retention Index
- This section allows the system to calculate the retention
index window in one of three ways. The retention index calibration data are a result of the
retention calibration setup in the Gas Chromatograph window of the Chromion software.
•
Retention index from file
uses the calibration information that is stored in the data
file.
•
Retention index from file
uses the calibration information that is stored in the data
file.
•
Manual calibration from method
uses the retention index calibrations manually
added in Gas Chromatograph. Refer to the Gas Chromatograph section of this chapter
for instructions on setting up manual retention index calibrations.
•
Auto calibration from method
uses the retention index calibrations appearing in Gas
Chromatograph.
Retention index
is the value appearing in the deconvolution window for the unknown
compound.
Range
This section allows manual entry of values for any of the fields.
•
RI-/RI+
If wider windows are needed the values can be changed on this screen
without changing the default settings in the method.
•
Window
The number of the Window is the plus and minus (+/-) of a value of the
retention index of the compound in which identification can be made of a compound.
The number is set the Window in Deconvolution Parameters of Target Library Editor for
the unknown library used when searching compounds. If the retention index is 475 then
Содержание TORION T-9
Страница 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Страница 4: ...iv Torion T 9 User s Guide...
Страница 10: ...6 Torion T 9 User s Guide...
Страница 11: ...1 Introduction...
Страница 15: ...2 Safety Practices...
Страница 24: ...20 Torion T 9 User s Guide...
Страница 31: ...3 System Overview...
Страница 40: ...36 Torion T 9 User s Guide...
Страница 41: ...4 Basic Custodion Operations...
Страница 49: ...5 Basic Operation...
Страница 116: ...112 Torion T 9 User s Guide...
Страница 117: ...6 Software...
Страница 257: ...7 Advanced Operations...
Страница 260: ...256 Torion T 9 User s Guide...
Страница 261: ...8 Maintenance...
Страница 268: ...264 Torion T 9 User s Guide...
Страница 269: ...9 Troubleshooting...