72 . Torion T-9 User’s Guide
CAUTION
ATTENTION
Be sure the face of the Custodion syringe (the red disc w ith the
gold rings) makes good contact w ith the contact points on the
instrument as soon as the syringe is inserted.
Assurez-vous que le visage de la seringue du Dépositaire (le
disque rouge avec les bagues en or) entre en contact avec les
points de contact de l'instrument dès que la seringue est
insérée.
The gold rings on the Custodion SPME syringe must contact the points on the GC injection
port. The instrument will NOT start an analysis if the Custodion syringe does not come in
contact with these contact points. The following figure shows (1)
SPME syringe gold rings (2)
GC injection port contacts.
The screen will display “
Please Wait…
” while the standard analytes are being desorbed from
the SPME.
After the specified time the screen will prompt the user to
Depress and release plunger,
then remove syringe
.
NOTE: It is important to be sure the plunger is not depressed until the SPME syringe is fully inserted.
However, the time between the insertion of the syringe and the depression of the plunger should be
minimal, so no vapors are lost during this sample introduction.
Содержание TORION T-9
Страница 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Страница 4: ...iv Torion T 9 User s Guide...
Страница 10: ...6 Torion T 9 User s Guide...
Страница 11: ...1 Introduction...
Страница 15: ...2 Safety Practices...
Страница 24: ...20 Torion T 9 User s Guide...
Страница 31: ...3 System Overview...
Страница 40: ...36 Torion T 9 User s Guide...
Страница 41: ...4 Basic Custodion Operations...
Страница 49: ...5 Basic Operation...
Страница 116: ...112 Torion T 9 User s Guide...
Страница 117: ...6 Software...
Страница 257: ...7 Advanced Operations...
Страница 260: ...256 Torion T 9 User s Guide...
Страница 261: ...8 Maintenance...
Страница 268: ...264 Torion T 9 User s Guide...
Страница 269: ...9 Troubleshooting...