Shutdown . 109
time. After the button is no longer active the venting process will continue and the shutdown
process cannot be aborted.
4.
Cancel Shutdown
is the second button that becomes available on the shutdown screen.
Pressing it
will stop the shutdown process and restore the pumping system and the heaters.
The home screen will display when the system is ready to run samples again.
Exiting the Softw are
The second shutdown process is used to exit the current software that is being used in order to access
additional software on the instrument. The pumps and heaters will be left running so the system can
quickly be brought back up again.
1.
From the home screen click on the
Advanced
button.
2.
Then click the
Status/Diagnostics
button.
3.
Go to the
Edit
tab of the Status/Diagnostics window and click on the
Exit Software
button.