42 . Torion T-9 User’s Guide
L iquid Unknown S ample
Take headspace of sample in environment
Increase exposure time
P lace sample into a vial and take headspace
Heat vial (warm with hand or other means
such as equipment in mobile lab) and take
headspace
Immerse S P ME directly into liquid
Take headspace with shorter exposure time
R educe the time of exposure in injector port.
Use a syringe to take the headspace and
deliver it into another vial. Take headspace
of the transferred headspace.
Allow sample on S P ME to evaporate before
injection (do this by exposing the S P ME
fiber to the atmosphere)
Dilute sample in appropriate solvent.
Immerse S P ME directly in liquid of the
dilution. Do this in a mobile lab under
engineering control. Be careful of possible
reactions with solvent. Use the M-8 paper
test to determine whether an aqueous or an
organic solvent should be used.
If sample detection is too low
If sample detection is too high
Содержание TORION T-9
Страница 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Страница 4: ...iv Torion T 9 User s Guide...
Страница 10: ...6 Torion T 9 User s Guide...
Страница 11: ...1 Introduction...
Страница 15: ...2 Safety Practices...
Страница 24: ...20 Torion T 9 User s Guide...
Страница 31: ...3 System Overview...
Страница 40: ...36 Torion T 9 User s Guide...
Страница 41: ...4 Basic Custodion Operations...
Страница 49: ...5 Basic Operation...
Страница 116: ...112 Torion T 9 User s Guide...
Страница 117: ...6 Software...
Страница 257: ...7 Advanced Operations...
Страница 260: ...256 Torion T 9 User s Guide...
Страница 261: ...8 Maintenance...
Страница 268: ...264 Torion T 9 User s Guide...
Страница 269: ...9 Troubleshooting...