AAnalyst 200 User’s Guide
3-7
Location and Space Requirements
Bench and location requirements
The location you choose for the spectrometer must fulfill the criteria listed below:
•
Place the spectrometer on a sturdy bench or trolley (cart) capable of sustaining the weight.
The work surface must be level, flat, clean, dry, and free from vibration. It should be strong
and stiff enough to bear the weight of the spectrometer system over a long period without
warping or bending. A movable table is very convenient, but make sure that the wheels can
be securely locked.
•
Position the bench or trolley out of direct sunlight and away from radiators and heaters.
•
Position the system near to the electricity and gas supply points.
•
Position the system so that the atomizer compartment is located beneath a suitable exhaust
venting system.
•
The instrument should not be against a wall, as the instrument must be accessible from the
back for servicing purposes. An accessible space of at least 25.4 cm (10 in.) should be
available behind the instrument.
•
Make sure that there is space at the rear and sides of the system for air to circulate freely. Do
not block the fan operating on the right side of the instrument.
•
Make sure that there is sufficient room for you to work comfortably with the instrument, and
that you can reach the connections at the rear of the instruments and the electricity and gas
supply points.
Laboratory Space Requirements
The illustrations on the following pages show the dimensions of the spectrometer and the space
that you require to accommodate the spectrometer, and ancillary equipment.
Instrument Dimensions
AAnalyst 200 is 700 mm wide (27.6 in.), 760-mm deep (29.9 in.), 650-mm (25.5 in.) high. The
instrument weighs 49 kg (108 pounds).
The main power cable length is 2 m (6 ft).
To provide access for servicing and space for ventilation, allow at least 25.4 cm (10 in.) of space
between the instrument and the laboratory walls.
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......