Recommendations for compressed gas systems
Gas installations
•
Legibly mark gas lines and outlets to clearly identify the contents. Use the chemical name or
commercially accepted name for the gas.
•
Use only approved regulators, tubing, and hose connectors.
•
Arrange gas hoses where they will not be damaged or stepped on and where things will not
be dropped on them.
•
Check the condition of pipes, hoses, and connectors regularly, and replace any damaged
parts.
•
Perform periodic gas leak tests at all joints and seals of the gas system by applying an
approved gas leak detection fluid.
Using gas cylinders
•
Store cylinders in accordance with the applicable national, state, and/or local regulations and
standards.
•
Legibly mark cylinders to clearly identify the contents and status—full or empty. Use the
chemical name or commercially accepted name for the gas.
•
When gas cylinders are stored in confined areas, such as a storage room, make sure that
ventilation is adequate to prevent toxic or explosive accumulations of gas.
•
Do not store acetylene in close proximity to oxidizing gases, such as nitrous oxide.
•
Store all gas cylinders upright, with the valve cap in place, in a dry, well ventilated area,
protected against temperature extremes.
•
Do not locate gas cylinders near exits and exit routes, elevators, gangways, or in locations
where heavy moving objects may strike or fall against them.
•
Locate gas cylinders away from heat sources, including the direct rays of the sun, and away
from ignition sources and readily ignitable substances, such as gasoline, oil, or combustible
waste. Compressed gas cylinders should not be subjected to temperatures above 52 °C (126
°F).
•
Locate all gas cylinders only in a vertical position, fastened securely to an immovable
bulkhead or a permanent wall.
•
Move cylinders with a suitable hand truck after making sure that the valve cap is securely in
place and that the cylinder is properly fastened to the hand truck.
•
Do not attempt to refill gas cylinders.
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......