AAnalyst 200 User’s Guide
9-45
Replacing Parts of Wear
Burner System O-Rings
There are o-rings in the neck of the burner chamber and in the burner end cap. You must check
the condition of these o-rings at regular intervals and replace them if they are damaged or show
signs of deterioration.
WARNING
Warning: Flammable Gases – Fire and Explosion Hazard
If the o-rings in the burner system are damaged or have deteriorated, they may
no longer form a proper seal. This may result in a gas leak which is a serious
fire hazard.
Check the condition of the o-rings regularly and replace them if they are
damaged or show signs of deterioration.
1.
Take the burner system apart.
2.
O-ring located in the neck of the burner chamber. Carefully lever the o-ring out of the neck of
the burner chamber using a plastic pipette tip or the like.
NOTE: Never use sharp or pointed tools to remove the o-ring. Examine the o-ring. If it is damaged,
replace it.
1.
Place the o-ring back into the neck of the burner chamber and press it down with your finger.
2.
O-ring located in the burner end cap. Very carefully roll out the o-ring with your finger.
3.
Examine the o-ring. If it is damaged, replace it.
4.
Gently place the o-ring back into the end cap and press it into place with your finger.
5.
Reassemble the burner system.
Nebulizer O-Rings
There are 3 o-rings on the nebulizer. Always carefully check the condition of these o-rings
whenever you take the nebulizer apart.
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......