Touch Screens
7-16
Items on this page
Integration Time
(sec)
Touch-entry field that displays the selected integration time. This is
the time period over which the spectrometer takes a measurement. The
longer the time, the greater is the level of signal integration.
To change the integration time, touch this field to display the number
pad dialog.
Enter the required integration time and then touch
OK
.
Replicates
Touch-entry field that displays the selected number of replicate
measurements. The selected number of measurements is performed on
each solution (blank, standard, and test sample).
To change the number of replicates, touch this field to display the
Replicates dialog. Enter the required number of replicates and then
touch
OK
.
Read Delay (sec)
Touch-entry field that displays the selected read delay. This is the time
period that the spectrometer waits before it starts to measure the
signal. A read delay is typically used to allow the flame to stabilize
after a change of solution.
To change the read delay, touch this field to display the Read Delay
dialog. Enter the required read delay and then touch
OK
.
This item allows you to select the information you would like to have
printed together with the analytical mean and statistical results. You
select whether printing should be on or off via the
Printer On/Off
button on Analyze page.
Sample Handling
When the (optional) autosampler is installed, you can choose between
manual data display, manual continuous graphics display, and
autosampler.
Measurement
(MHS only)
Peak Height
measures the largest signal during the Read Time after
the selected smoothing is applied.
Peak Area
measures the area under
a peak in absorbance-seconds.
Smoothing Points
(MHS only)
Smoothing Points uses the Savitsky-Golay smoothing algorithm to a
group of
n
data points; where
n
can be 0, 5, 9, 19, and 37. Smoothing
is applied to the geographical display and used in the calculation of the
Peak Height, when the Peak Height measurement mode is selected.
Peak Plot Scale
(MHS only)
Automatic
sets the scale based on the highest point on the plot being
displayed.
Fixed absorbance
is manually setting peak scales.
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......