AAnalyst 200 User’s Guide
9-13
Cleaning the Burner Chamber
NOTE: Do not scrape the plastic components with metal tools. Do not soak the burner chamber in acid
or use abrasive cleaner.
1.
Remove the burner system from the atomizer compartment and take the burner chamber
apart.
2.
Clean the burner chamber and flow spoiler with a soft brush and a mild laboratory detergent
solution.
3.
Clean the inside of the burner end cap with a soft brush and a mild laboratory detergent
solution.
4.
After cleaning, rinse thoroughly with deionized water.
5.
Reassemble the burner system.
Cleaning the burner chamber after use with organic solvents
If you aspirate aqueous samples immediately after organic samples (e.g. oils or MIBK extracts),
the absorption signals can be noisy and erratic. After aspirating organic samples, perform the
procedure below to prevent contamination of subsequent aqueous samples.
1.
Aspirate for 5 minutes an organic solvent that is miscible with the samples that have just been
aspirated.
2.
Aspirate methanol or another solvent miscible with both the organic solvent and water, for 5
minutes.
3.
Aspirate 1% nitric acid solution for 5 minutes.
4.
Flush the drain tube thoroughly with water. Empty the collection vessel and refill with water.
Dispose of hazardous or corrosive solutions properly and observe local codes concerning the
effect of the waste on the environment.
NOTE: Dispose of hazardous or corrosive solutions properly and refer to your local safety regulations
for proper disposal procedures.
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......