Analyzing Samples
8-16
Analyzing Samples Using Linear Calibration
The following example describes how to run samples using the linear calibration technique. It
should be noted that the same steps are carried out when using the nonlinear calibration
technique. The only difference is in setting up the calibration type. Instead of selecting nonlinear
for the calibration type, you select nonlinear.
EXAMPLE:
Determining copper using linear calibration
The procedures in this example assume that you have performed the set-up procedures in
Chapters 3, 4, 5, and 6, in which case the instrument will be on and the flame ignited.
This example runs a blank, 3 standards, and 3 test samples. We instructed the autosampler to
perform a reslope after the second test sample.
1.
Carry out the following steps to duplicate our example:
2.
Prepare solutions. Set up the autosampler
3.
Load the autosampler
4.
Enter parameters
5.
Start the run
Detailed procedures follow.
Prepare solutions
In this example, we used a copper lamp and solutions of copper for our determinations. However,
you do not have to use copper. For your convenience in making substitutions we have included a
table of recommended concentrations for Ag, Mg, Mn, and Pb.
1.
Prepare about 200-250 mL of 1% v/v HNO
3
to use as a blank solution.
2.
Prepare about 200-250 mL each of three standards in 1% v/v HNO
3
of the element you want
to determine. Suggested standard concentrations for five elements are shown in the following
table:
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......