Safety Practices
2-14
Compressed Gases
NOTE: The permanent installation of gas supplies is the responsibility of the user and should conform to
local safety and building regulations.
WARNING
Warning: Compressed Gases
High pressure gas cylinders can be dangerous if mishandled or misused. You
must have a tank regulator on the gas cylinder.
NOTE: The responsible body must ensure that the type of connector used at the outlet side of the gas-
pressure regulator conforms to applicable national requirements.
Summary of gas hazards
Hazards associated with the different gases used in atomic absorption spectrometry are presented
in the table below:
Gas Suffocation
Explosion
Spontaneous
Decomposition
or Combustion
Air
Argon x
Nitrogen x
Acetylene (ethyne, C
2
H
2
) x x
x
Nitrous Oxide (dinitrogen
monoxide, N
2
O)
x
x
Contact the gas supplier for a material safety data sheet (MSDS) containing detailed information
on the potential hazards associated with the gas.
WARNING
Warning: Nitrous oxide is a strong oxidant and an asphyxiant
Nitrous oxide is an asphyxiant; it can be considered a potential hazard whenever it is stored or used in a
confined space.
Spontaneous combustion may occur if nitrous oxide comes into contact with grease, oil or other organic
materials.
•
Store nitrous oxide cylinders in a well-ventilated area outside the laboratory.
•
Do not store nitrous oxide cylinders in close proximity to flammable gases (e.g. acetylene).
•
Make sure tubing and fittings carry nitrous oxide are free of grease, oil and other organic material.
WARNING
Warning: Acetylene-Explosion Hazard
Acetylene can decompose explosively at pressure higher than 103 kPa/1.03 bar/15 psig.
•
Always make sure that the acetylene outlet gauge pressure is below these values.
Acetylene can react with copper to form a readily explosive compound.
•
Never use copper tubing or fittings for acetylene lines.
Make sure that all gas connections are gas-tight. Check the connections and gas hoses regularly
for leaks using a leak testing fluid.
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......