AAnalyst 200 User’s Guide
8-17
Table 8- 2 Suggested Standards
Element Standard
1
(mg/L)
Standard 2
(mg/L)
Standard 3
(mg/L)
Reslope
(mg/L)
Ag 0.2
0.5 1 0.5
Cu 0.5 1 1.5 1
Mg 0.05
0.1 0.2 0.1
Mn 0.2
0.4
0.6 0.4
Pb 2 5
10 5
3.
For samples 1, 2 and 3, use Standards 1, 2, and 3, respectively.
Set up the autosampler
The autosampler you intend to use must be connected to the spectrometer. Installation procedures
are described in the
User’s Guide
provided with the autosampler.
Set up the autosampler by following the directions given in Chapter 6
, Operation of the
Instrument.
Load the autosampler
In the autosampler tray, location zero (0) is reserved for the wash, (1) is for blank, and locations 2
through 9 are reserved for standard solutions. The last standard location is reserved for reslope.
If not already done, load the autosampler tray with a blank, standards, and samples.
Enter parameters
Refer to Chapter
Touch Screen
, section
Analyze Page: Autosampler
, if you require information
on how to enter analytical parameters. Fill in the entries for autosampler tray, first and last sample
location, and recalibration frequency.
If not already done, install and align the required lamp, and ignite the flame.
On Parameters page, Spectrometer subpage:
1.
Enter 1 second for the
Integration Time
.
2.
Enter 3 for the number of
Replicates
.
3.
Enter 3 seconds for the
Read Delay
.
4.
Enter the required
parameters.
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......