AAnalyst 200 User’s Guide
6-31
5.
Touch the On/Off check box. If necessary, change the lamp current. Touch
OK
.
6.
Touch the
Element
entry field box to select the desired element. Touch
OK
.
7.
Touch the
Setup Instrument
button on the Lamp page of the touch screen. The recommended
parameters for that element are automatically set up.
•
The system automatically switches the lamp on, sets up the optical path (prism and grating),
sets and peaks the wavelength, sets the slit, and performs an Automatic Gain Control (AGC)
and Automatic Intensity Control (AIC).
8.
Turn the two alignment controls alternately and in small increments while watching the AA bar
graph and the numerical energy display. Align the lamp to obtain maximum energy.
Uncoded multi-element Lumina lamp
1.
Follow steps 1-5 of the coded single element Lumina lamp.
2.
Touch the
Element
entry field box on the Lamp page to display the Select Element dialog. Select
the desired elements. Use the
Clear
button to remove unwanted element from the selected list.
Touch
OK
.
3.
Touch the
Setup Instrument
button on the Lamp page of the touch screen. The recommended
parameters for that element are automatically set up.
•
The system automatically switches the lamp on, sets up the optical path (prism and grating),
sets and peaks the wavelength, sets the slit, and performs an Automatic Gain Control (AGC)
and Automatic Intensity Control (AIC).
4.
Turn the two alignment controls alternately and in small increments while watching the AA bar
graph and the numerical energy readout on Lamp page. Align the lamp to obtain maximum energy.
Turret coded Lumina lamp (mount)
1.
Start up the spectrometer, but do not ignite the flame.
2.
If the desired element already appears in the Element entry field box, then just touch the
Setup
Instrument
button.
3.
If you like to select a different element, touch the
Element
entry field box on the Lamp page to
display the Element dialog. Select the desired element and touch
OK
.
•
The system automatically switches the lamp on, sets up the turret and the optical path (prism
and grating), sets and peaks the wavelength, sets the slit, and performs an Automatic Gain
Control (AGC) and Automatic Intensity Control (AIC).
Содержание aanalyst 200
Страница 1: ...AAnalyst 200 User s Guide ...
Страница 6: ......
Страница 7: ...Introduction 1 ...
Страница 8: ......
Страница 18: ...Introduction 1 12 ...
Страница 19: ...Safety Practices 2 ...
Страница 20: ......
Страница 44: ...Safety Practices 2 26 ...
Страница 45: ...Preparing Your Lab 3 ...
Страница 46: ......
Страница 52: ...Preparing Your Lab 3 8 473 mm 650 mm 910 mm 700 mm 760 mm 170 mm Figure 3 1 Outside dimensions of instrument ...
Страница 55: ...AAnalyst 200 User s Guide 3 11 Figure 3 3 Space requirements for the spectrometer system top view ...
Страница 66: ...Preparing Your Lab 3 22 ...
Страница 67: ...System Description 4 ...
Страница 68: ......
Страница 89: ...Installation 5 ...
Страница 90: ......
Страница 107: ...AAnalyst 200 User s Guide 5 19 Setting up the Burner System Figure 5 9 Full view of the AAnalyst 200 Instrument ...
Страница 115: ...Operation of the Instrument 6 ...
Страница 116: ......
Страница 126: ...Operation of the Instrument 6 12 Installing the Burner Mount Assembly Figure 6 6 Installing the burner mount assembly ...
Страница 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Страница 158: ...Operation of the Instrument 6 44 ...
Страница 159: ...Touch Screens 7 ...
Страница 160: ......
Страница 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Страница 193: ...AAnalyst 200 User s Guide 7 35 Screen 7 22 Shortcut menu Properties folder appear and select Share this folder option ...
Страница 210: ...Touch Screens 7 52 ...
Страница 211: ...Analyzing Samples 8 ...
Страница 212: ......
Страница 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Страница 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Страница 244: ...Analyzing Samples 8 34 ...
Страница 245: ...Maintenance 9 ...
Страница 246: ......
Страница 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Страница 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Страница 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Страница 299: ...AAnalyst 200 User s Guide 9 55 Figure 9 20 The centering of the two lamp beams The alignment procedure is now complete ...
Страница 307: ...Troubleshooting 10 ...
Страница 308: ......
Страница 320: ...Troubleshooting 10 14 ...
Страница 321: ...Index ...
Страница 322: ......