ESO 2 ICU Ventilator
61
User Manual
Modes Menu
ESCAPE FROM MENU
SUPPORT MODE: SIMV, SMMV,
PSV
VOLUME TYPE: Tidal
SIGH ENABLE:
SIGH TO BREATH RATIO:
INSP. PAUSE% : 0%
APPLY: SITE DEFAULT
In SPONT mode and VOLUME mode, and SIMV/ SMMV, the MODES
menu (accessed by touching the mode icon area on the screen) is:
ESCAPE FROM MENU
VOLUME TYPE: Tidal
SIGH ENABLE:
SIGH TO BREATH RATIO:
(2)
INSP. PAUSE% : 0%
(3)
In PRESSURE mode and PSV modes the MODES menu (accessed by
touching the mode icon area on the screen) is:
ESCAPE FROM MENU
SIGH ENABLE:
SIGH TO BREATH RATIO:
(2)
INSP. PAUSE% : 0%
(3)
Notes
1) Support mode depends on configuration options.
The SUPPORT MODE option will be missing from the MODE menu if:
a) Options are not enabled
b) ‘SPIROMETRY: off’ is displayed.
The support mode sub menu can include:
none / PSV / PSV, SIMV, SMMV
2) The options here are:
SIGH ENABLE: on - off
SIGH TO BREATH RATIO: 1:10 to 1:100
(1:10 indicates 1 breath with sigh, then 10 breaths without sigh)
3) The options here are: 0 - 60%
Appendix
Service Menu
ESCAPE FROM MENU
LANGUAGE: ENGLISH
PATIENT LOG MENU
*SITE DEFAULTS
SERIAL MODE: none
ABSORBER SWITCH: ON
UPGRADE MENU
AMBIENT PRESSURE: 988 mbar
DISPLAY HISTORY
**SERVICE PIN: 0
**ENGINEER MENU
NOTES
*The values listed for site defaults are examples
**Sub-menus for Service PIN and Engineer Menu are not
accessible by users.
Upgrade Menu
ESCAPE FROM MENU
I/O FIRMWARE: vx.xx [Build xx]
MAIN FIRMWARE: vx.xx [Build xx]
REGISTRATION KEY: unknown
UPGRADE FIRMWARE: unavailable
ADD NEW FEATURE: unavailable
DEVICE TYPE: AVS
SERIAL NUMBER: xx xx xx
LOCK STATE: on
Display History
ESCAPE FROM MENU
MANUFACTURER DATE : 03/03/05
TOTAL HOURS RUN: 100
LAST SERVICE DATE: 13/08/04
HOURS SINCE SERVICE: 100
DRIVE VALVE CYCLES: 1253
PATIENT VALVE CYCLES: 822
CUTOFF VALVE CYCLES: 72
Patient Log Menu
ESCAPE FROM MENU
PRINT PATIENT DATA
LOGGING: off
LOG STATUS: disabled
CLEAR LOG DATA
LOGGING WINDOW: 10 min
Site Defaults
ESCAPE FROM MENU
SAVE TO SITE
VIEW: SITE DEFAULTS
VOLUME TYPE : tidal
V
T
SET: 550 ml
V
M
SET: 5.5 Litres
T+PS INIT: 10 cmH2O
SET BPM : 10
I : E : 1:1.0
PEEP : OFF
LIMIT : 38 cmH2O
TRIGGER : 10 L/min
APNOEA ALARM LIMIT : 15 sec
The MODES menu is context sensitive, with content
dependent on current mode.
In STANDBY the MODES menu is:
ESCAPE FROM MENU
SUPPORT MODE: SIMV, SMMV, PSV
(1)
VOLUME TYPE: Tidal
SIGH ENABLE:
SIGH TO BREATH RATIO:
(2)
INSP. PAUSE% : 0%
(3)
APPLY: SITE DEFAULT
Содержание ESO 2
Страница 1: ...ESO2 Emergency Ventilator User Manual VENTILATION SOLUTIONS...
Страница 2: ......
Страница 60: ...ESO 2 ICU Ventilator 54 User Manual...