ESO 2 ICU Ventilator
14
User Manual
Description
3.2.8.1 Fresh Gas Compensation
Adjusts delivered volume up to 60%
An alarm is triggered if measured volume varies by 50% from
the set volume.
This function is user adjustable (3.2.9).
NOTE
Fresh gas compensation is disabled if:
The spirometry system is turned OFF through the menu
system, or the spirometry system is malfunctioning.
3.2.8.2 Fresh Gas Mixture Compensation - models with
Spirometry
The spirometry system compensates for the fresh gas mixture
in use.
The Gas Mixture window (Area C) is an active touch-selectable
area (in any mode), with a drop down menu.
1. Touch the screen in Area C.
2. Use the wheel to select:
O2 + AIR
Do
not
select O2 + N2O
3. Press to confirm. or, use the menu structure, as shown.
Select the required mixture with the scroll wheel.
NOTE
1. Fresh gas mixture compensation is disabled if:
The spirometry system is turned OFF through the menu
system, or the spirometry system is malfunctioning.
2. If the oxygen monitor is switched OFF, a default mixture is
assumed.
3.2.8.3 Compliance Compensation
The ventilator will apply compliance compensation to account
for compliance loss in the breathing system in cases where:
Fresh gas compensation is disabled, or
Spirometry is unavailable or disabled
NOTE
For correct operation the value of the breathing system
compliance must be established first, by completing the
ventilator leak-test as part of the Pre-operation Procedure.
Refer to section 4, noting that breathing system compliance is
displayed as ‘Bsys.comp’
If the leak test is not carried out, the default value will be used
(7 ml/cmH2O).
NOTE
In compliance compensation mode any fresh gas used will be
in addition to the set tidal volume.
3.2.8.4 Altitude Compensation
This function monitors ambient pressure, and adjusts the
delivered volume accordingly
Altitude compensation is automatically applied during
calibration of the oxygen monitor - see section 4.3
.
Area A
Area C
Area D
Area B
Area E
Содержание ESO 2
Страница 1: ...ESO2 Emergency Ventilator User Manual VENTILATION SOLUTIONS...
Страница 2: ......
Страница 60: ...ESO 2 ICU Ventilator 54 User Manual...
Страница 71: ......