Description
3.2 Ventilator
3.2.1 Description
• The AV-S Ventilator is a pneumatically driven, software
controlled, multi-mode ventilator.
• The ventilator is time-cycled, volume/pressure controlled,
and pressure limited.
• The ventilator has compliance compensation and fresh
gas compensation.
• User-selectable gas mixture compensation is a standard
feature, plus a user-selectable variable inspiratory pause
and sigh option.
Ventilation Modes
• Volume Mode:
continuous mandatory ventilation
• Pressure Mode:
pressure controlled mandatory ventilation
• PEEP:
available in volume and pressure modes
• Spontaneous and advanced patient support:
SIMV, SMMV,
PSV.
See Appendix
Patient Monitoring
• Airway pressure, measured from the expiratory limb of the
breathing circuit.
• Tidal Volume and Minute Volume display is provided by a
dual spirometry system.
• An integral oxygen monitor system measures oxygen
concentration in the breathing circuit inspiratory limb.
• The print function provides a permanent record of
ventilator function activity for up to eight hours procedure,
or can be used to record waveforms.
Drive gas supply
WARNING
1. The ventilator is designed to be driven by medical air
only.
The drive gas is set during manufacture and the
ventilator is calibrated for that gas.
Before the ventilator is used clinically for the first time,
the commissioning engineer must confirm that the drive
gas setting is set correctly to Air.
The use of any other gas will cause inaccurate operation
and may damage the ventilator, resulting in potential
injury to the patient.
1. The driving gas supply must be clean and dry to prevent
ventilator malfunction.
• Supply pressure: Air at 280 to 600 kPa (41 to 87 psi).