ESO 2 ICU Ventilator
17
User Manual
Description
3.2.9.5 Volume Type Selection
Touch the screen tab to switch between Tidal Volume and
Minute Volume.
NOTE: Minute Volume is derived from a rolling average during
a 30 second period.
3.2.9.6 Volume Mode Operating Functions
1. Inspiratory Pause function:
Inspiratory pause can be varied in the menu from 0 - 60%.
The inspiratory pause menu can also be accessed by
touching the icon area of the screen.
WARNING : This can affect the maximum Tidal Volume.
Select Inspiratory Pause:
• Press the menu switch (1), turn the wheel (2) to scroll
through the menus
• Select Modes
• Select percentage time value (0 - 60%)
• Exit menus
The symbol for inspiratory pause will
appear on the display
NOTE:
The inspiratory pause function is cancelled when
Standby is selected
2. Sigh function:
• ‘Sigh’ option is available in Volume Cycle mode.
Sigh to breath ratio is 1:n, where n has a range of 10 - 100.
The Sigh menu can also be accessed by touching the icon
area of the screen.
NOTE
a) A sigh to breath ratio of 1:10 indicates 1 breath with
sigh, then 10 breaths without sigh
The extra volume will be approximately 50% above the
tidal volume set by the user (1.5 x set V
T
).
b) The High Volume Alarm is not triggered when ‘Sigh’ is
selected.
•
Select Sigh function:
• Press the Menu switch (1), turn the wheel (2) to scroll
through the menus
• Select Modes
• Select Sigh Enable on/off
• Select Sigh-to-Breath Ratio
• Rotate the wheel to select the required value
• Press wheel to confirm, and exit menus
The legend for Sigh will appear on the
display
NOTE:
The sigh function is cancelled when standby is selected.
EXIT MENUS
O2 MONITOR & SPIROMETRY
LEAK TEST
FRESH GAS COMPENSATION:ON
MODES
WAVEFORM
ALARM SETTINGS
GAS MIXTURE: O2+AIR
SERVICE MENU
2
1
3.
Volume measurement
Volumes are measured if the Spirometry function is
selected.
Automatic High or Low volume alarms are triggered if the
measured volume is 50% above or below the set volume.
4.
User adjustable option
If the maximum pressure limit is achieved, the ventilator
cycles to the expiratory phase.
3.9.2.5 Touchscreen Access to Mode Configuration Options
Touch the screen in the area containing the green icons to
access mode configuration options (including INSP PAUSE,
SIGH, and APNOEA ALARM mute/inhibit).
V
M
EAS
V
T
MEAS