ESO 2 ICU Ventilator
27
User Manual
Pre-operation Procedures
1. Turn the flow off at the Gas Delivery on/off switch. This will stop
all gas flows (including the basal flow).
This will also turn the ventilator off.
2. Turn the ventilator switch on (1) (do not use the gas delivery
switch).
3. Remove the breathing circuit hoses from the inspiratory and
expiratory connectors on the breathing module.
4. Ensure that the ventilator bellows is empty,
5. Calibrate the spirometer via the ventilator menu procedure.
6. Press the menu switch (2) on the front panel.
7. Scroll down the main menu and select O2 MONITOR &
SPIROMETRY.
8. Select SPIRO CALIBRATION.
9. Press the wheel (3) to initiate calibration.
10. Calibration is completed.
11. Scroll to ESCAPE FROM MENU.
12. Press the wheel to confirm.
4
.1.7 Pressure Monitor Connections
WARNING: The High and Low Airway Pressure Alarms are
important for patient care.
1. Check the pressure monitor tubing assembly connections (1).
2. Push-fit, self-sealing connectors (2):
• Push in the tube as far as possible
• Do not use excessive force.
• The connector end piece ‘X’ will also move inwards.
• Pull the tube carefully outwards. The end piece ‘X’ will be pulled
outwards to the ‘locked’ position.
Tubing disconnection or sensor malfunction
Message displayed: ‘Check pressure sensing’
Action by user:
1. Check the condition of the tubing, and the connections (1) at the
ventilator and the rear of the breathing module .
2. If the tubing is undamaged and the connections are secure, the
operation of the sensors must be checked by a service engineer
CAUTION
The ventilator will apply compliance compensation to account for
compliance loss in the breathing system in cases where:
1. Fresh gas compensation is disabled, or
2. Spirometry is unavailable or disabled
The ventilator will continue to function, although the target pressure
may be exceeded by up to 10 cmH2O.
X
X
2
1
1
O2 Monitor & Spiro
ESCAPE FROM MENU
O2 MONITOR: on
CALIBRATION: 100%
HIGH ALARM SET: 105
LOW ALARM SET: 18
SPIROMETER: on
>
SPIRO CALIBRATION: 0 L/min
3
2
1
Содержание ESO 2
Страница 1: ...ESO2 Emergency Ventilator User Manual VENTILATION SOLUTIONS...
Страница 2: ......
Страница 60: ...ESO 2 ICU Ventilator 54 User Manual...
Страница 71: ......