ESO 2 ICU Ventilator
55
User Manual
Appendix
APPENDIX 6 Specification
Application
Ventilation of adult and paediatric patients
Internal Compliance
Adult bellows
Paediatric bellows
3 ml/cmH2O (nominal)
2 ml/cmH2O (nominal)
Physical
Size (mm)
- control unit only
- with adult bellows
290 wide x 300 deep x 185 high
290 wide x 300 deep x 385 high
Screen Size
(a) 210 mm (8.4 inches) TFT
(b) 305 mm (12 inches) TFT
Weight - control unit only
- with adult bellows
7.6 kg
9 kg
Bellows
Adult
Paediatric
(Note - bellows are latex free)
20 - 1600 ml
20 - 350 ml
Electrical Power
110 - 120 VAC, 50 - 60 Hz, 1.2 A max
200 - 240 VAC, 50 - 60 Hz, 0.6 A max
(automatic ranging)
Battery Back-up
12 V, 2 Ah, sealed lead-acid battery
Fully charged battery provides one hour (nominal) backup
Fuse (mains supply)
Type T 2AH, fused L and N
2 A, 250 V rating, 20 mm, anti surge, ceramic
Fuse (battery)
3 A fastblow, 32 V rating, 1000 A interrupt, mini-blade type
10.9 x 8.8 x 3.8mm
Drive Gas
Oxygen or Air (dry, and oil free) at 280 to 600 kPa (41 to 87 psi)
Alarms
Alarm Mute
30 or 120 seconds (see 3.12)
Apnoea
Flow referenced (no breath detected within 15 seconds)
High Continuous Airway Pressure
PEEP + 10 cmH2O for 1 cycle, or PEEP + 5 cmH2O for 3 cycles
A PEEP value of 4 cmH2O is used for PEEP values ≤4 cmH2O
Standby Mode: the alarm triggers after 15 seconds
Low Pressure (Pressure Control)
Bellows drive gas pressure sensor fails to detect at least 75% of set target level
Incorrect Rate or Ratio
Ventilator parameter settings unachievable
Mains Failure
Fully charged battery provides one hour (nominal) backup
Low Battery
If the ventilator is running on battery power, less than 20 minutes battery life
remains (see also, section 3.12)
Ventilator Inoperative
Internal failure
Outlet Blocked
Exhaust valve outlet blocked
Alarms - User Adjustable
Low Tidal Volume
Measured value is below 50% of volume set. Range: 10 - 1600 ml
High Tidal Volume
Measured value exceeds 150% of volume set. Range: 20 - 2400 ml
Low Minute Volume
Calculated value is 50% below volume set. Range: 0 - 50 L
High Minute Volume
Calculated value exceeds 150% of volume set. Range: 1 - 75 L
Low and High O2 Concentration
18% - 105%
High Airway Pressure
10 - 80 cmH2O adjustable