ESO 2 ICU Ventilator
42
User Manual
6.7 Sterilisation and Disinfectant Treatment Table
NOTE
1. After chemical disinfection, always rinse with warm water and dry in air.
2. Do NOT clean any component in an automatic cleaning/washing machine.
3. Before clinical use, ALWAYS carry out the Pre-use Checks listed in section 6 of this manual.
4. Refer to the instructions for use for the autoclave equipment.
Absorber
Component
Soap
water
Cidex sonacid
(see NOTE 1
above)
Steam
autoclave
Maximum temperature
°F
°C
Breathing hoses (check
manufacturer’s instructions)
Yes
Yes
Yes
278
137
Sa
f
elock fittings
Yes
Yes
Yes
278
137
Manifold block (including non-return
valves) and frame assembly
Yes
Yes
Yes
255
124
PEEP valve
Yes
Yes
Yes
278
137
Electrical interface unit
No
No
No
-
-
NOTE: During autoclaving, slight discolouration of the manifold block may occur.
This is not an indication of component deterioration.
Ventilator Bellows
Item
Method
Bellows
Gas, liquid, autoclave (20 cycles maximum)
Hoses
Gas, liquid, autoclave
O-rings
Gas, liquid, autoclave
Bellows base
Gas, liquid, autoclave
Exhalation valve
Gas, liquid, pasteurise, low temperature autoclave
Bellows canister
Liquid, autoclave
Ventilator Bellows - Reassemble
CAUTION
After cleaning, check that the small O-ring (1) located in
the bellows base under the diaphragm valve is in place. The
ventilator will not function if the O-ring is missing.
1. Refit the exhalation diaphragm valve assembly (2) to the
bellows base.
2. Refit the bellows assembly (3) and housing (4).
CAUTION
Check for correct fitment of the bellows (3). Function test the
ventilator before clinical use (refer to 4.2.1.3).
1
3
3
4
2
Maintenance