ESO 2 ICU Ventilator
44
User Manual
Appendix
APPENDIX 4 Operating system
Ventilator Control Unit Rear Panel
Gas Connections
1. Ventilator drive gas inlet
- connect to medical air supply
2. Bellows drive gas output
- connect to bellows via breathing module - see 5.1
3. Outlet - exhaled gas.
The discharged gas may contaminate
the environment. Connect to an AGSS system if available.
Electrical Connection
4. Electrical mains input and fuse unit
Interface and Parameter Inputs
5. Spirometer connection
6. Interface connector (not used)
7. Pressure monitor port
8. Input socket - oxygen monitor sensor
Data and Printer Ports
9. Data output
10. Output to ventilator remote display
11. VGA
12. Printer port
WARNING
The data output, VGA, and printer ports must only be
connected to equipment complying with IEC safety standards
(e.g. IEC 60950) which are then configured to comply with
IEC 60601-1 (Medical Systems).
7
8
6
9
10
11
12
4
2
1
3
5