ESO 2 ICU Ventilator
36
User Manual
5.4
Oxygen monitor
Switch on:
Select menus (1), select ‘O2 monitor & Spirometer’, select ‘O2
MONITOR: on’.
Use a bacterial filter in the breathing circuit expiratory limb to
protect the O2 sensor and breathing system.
Set High O2 Alarm
1. Access the sub menu and select ‘O2 MONITOR: on’.
2. Touch the O2 concentration display (2), or press the menu
switch (1) to select the O2 monitor sub-menu.
3. Scroll to HIGH ALARM SET, and press the navigator wheel
(3).
4. Rotate the wheel to change the displayed setting to the
desired value, press the wheel to confirm.
NOTE: The high alarm value (4) cannot be set below 19%
or above 105% (note that in certain conditions of excess
pressure, the readout may show a value above 100%).
Set Low O2 Alarm
The alarm value (5) cannot be set below 18%, or above 99%.
1. Select the O2 monitor sub-menu.
2. Scroll to LOW ALARM SET and press the wheel.
3. Rotate the wheel (3) to change the displayed setting (5) to
the desired value, press the wheel to confirm.
Select ESCAPE FROM MENUS, press the wheel to exit.
O2 Monitor & Spiro
ESCAPE FROM MENU
>
O2 MONITOR: on
CALIBRATION: 100%
HIGH ALARM SET: 105
LOW ALARM SET: 30
SPIROMETER: on
SPIRO CALIBRATION: 0 L/min
O2 Monitor & Spiro
ESCAPE FROM MENU O2 MONITOR: on
> CALIBRATION: 21%
HIGH ALARM SET: 105
LOW ALARM SET: 30
SPIROMETER: on
SPIRO CALIBRATION: 0 L/min
3
%O2
21
30
100
5
4
1
2
Start-up and operating instructions