6.4
Oxygen Sensor Replacement
These operations must be carried out by suitably trained
technicians only.
WARNING
The sensor contains:
1. A small quantity of electrolyte, classified as a harmful
irritant which is potentially hazardous.
2. Lead
Do not attempt to open a cell.
ALWAYS check the integrity of the sensor assembly before
use.
Once exhausted, the sensor must be disposed of according to
hospital, local, state and federal regulations.
6.4.1 Sensor Unit - Remove and Refit
Replacement sensor: Part No. 102714
1. Detach the cable connector (1) from the sensor (2).
2. Unscrew the sensor from the A200SP Absorber, and
discard. Dispose of the used components according to
hospital, local, state and federal regulations.
3. Screw the new sensor (2) into the absorber.
4. Attach the cable connector (1).
5. Calibrate the new sensor - see section 4.3.2.