ESO 2 ICU Ventilator
29
User Manual
Pre-operation Procedures
4.1.9 Bellows Assemblies
CAUTION
Always ensure correct fitment of bellows (see illustration
above), and carry out a full function test before clinical use, if
a bellows is removed and refitted.
Remove and refit
1. Remove the bellows housing (1).
Twist carefully counterclockwise until the bayonet tabs
become free, then lift up from the base (2).
2. Remove the bellows (3).
3. Refit the bellows and check for correct assembly, as
illustrated (4).
4. Fit the bellows housing (1) by pushing down, then twisting
clockwise until the bayonet tabs completely engage.
5. Function test the ventilator - section 4.2.1.3.
NOTE
If there is any malfunction, the ventilator must NOT be used.
If the problem cannot be rectified, the ventilator must be
checked by an engineer trained by the manufacturer.
Fitting a Paediatric Bellows Assembly
1. Remove the adult bellows housing (1) - twist carefully
counterclockwise until the bayonet tabs become free, then
lift up from the base (2). Remove the bellows (3).
2. Fit the paediatric adaptor (5). Press the adaptor firmly into
the ventilator bellows assembly base (2).
3. Fit the paediatric bellows (6) to the adaptor.
Check for correct assembly, as illustrated (4).
4. Fit the paediatric bellows housing (7) to the base by
pushing down, then twisting clockwise until the bayonet
tabs completely engage.
5. Function test the ventilator - section 4.2.1.3.
2
3
1
4
7
5
6