ESO 2 ICU Ventilator
15
User Manual
Description
3.2.9 Ventilation Modes
3.2.9.1 Standby Mode
1. Allows parameters to be set.
2. Some patient alarms are active:
• High airway pressure (at 80 cmH2O)
• High/Low oxygen
• Negative pressure
• Incorrect Rate/Ratio
• Continuous high pressure
• Target and pressure support
3. Access to Support modes is available in Standby mode
(depending on the support mode options on the ventilator):
• PSV
• SIMV
• SMMV
• SIGH ENABLE
SIGH TO BREATH RATIO
• INSP PAUSE
INSP PAUSE %
WARNING
PSV, SIMV, and SMMV are only available when spirometry
is enabled.
Note that patient support specifications vary. Support
modes are enabled at the factory, to the specification
chosen by the customer.
4. Displayed information
• Standby mode at ventilator start-up:
The last used Volume mode settings will be displayed.
• Standby mode selected while the ventilator is in use:
The screen will display the previous ventilation mode,
highlighted in yellow, within the relevant box.
The last used parameters will also be displayed.
• In Standby mode, the underline (1) indicates the last used
ventilation mode
1