ESO 2 ICU Ventilator
16
User Manual
Description
3.2.9.2 Volume Mode
WARNING
Important: Delivering high fresh gas flows when using closed
endotracheal suction.
This ventilator is configured to deliver flows of oxygen in excess
of the maximum shown on the flowmeter tube in order to fill the
ventilator bellows during closed endotracheal suction. Up to
approximately 30 L/min can be delivered.
Delivering O2 flows in excess of the maximum calibrated flow
will temporarily negate the possibility of accurate volume
delivery and measurement. Once the ventilator bellows is full
the O2 flow should be reduced to that in range of the calibrated
flow tube, then accurate volume delivery and measurement will
resume.
See section 5.3 for additional information.
Volume mode
The ventilator delivers a mandatory set volume of gas at
preset, fixed breath intervals.
The patient is making no respiratory effort.
3.2.9.3 Fresh Gas Compensation
The delivered volume is adjusted by up to 60%.
This delivered volume will consist of the volume delivered
from the ventilator bellows, plus the fresh gas flow from the
anaesthetic machine fresh gas supply, minus any compliance
loss and minus any leak.
This gives a total actual inspired tidal volume.
An alarm is triggered if the measured volume is 50% above or
below the set volume.
This function is user adjustable.
Compliance Compensation:
Refer to section 3.2.8.2
Altitude Compensation:
Refer to section 3.2.8.3
3.2.9.4 Select Volume Mode
Volume Mode selected from Standby Mode:
1. Press the screen tab: ‘VOLUME CONTROL’
Volume Mode selected from Pressure Mode:
1. Press the screen tab: ‘VOLUME CONTROL’ The ventilator
continues to ventilate in Pressure Mode.
2. The Volume Set display shows the previous setting, or
default setting.
3. A new Volume value can be set if required.
WARNING: Set appropriate values for the clinical procedure
in progress. Take note of all on-screen symbols and display
messages.
4. Press to confirm change of mode and new setting.
NOTE: Pressure limit will default to the previous Pressure
Target value + 5 cmH2O
5. At confirmation, the ventilator will switch to Volume Mode.
NOTE: Volume mode will commence at the beginning of an
exhalation phase.
Volume Mode
Parameters
Tidal volume
20 - 1600 mL
Rate
4 - 100 bpm
I:E ratio
1:0.2 to 1:8
PEEP ‘Off’ or
adjustable
4 - 20 cmH2O
(option, 4 - 30 cmH2O)
Pressure limit
10 to 80 cmH2O
To set, touch the ‘LIMIT’ screen
tab area, turn the navigator
wheel to select the required
value, press to confirm.
Inspiratory pause
Variable: 0 - 60% of inspiratory
(does not affect I:E ratio) time
Sigh
Approximately 1.5 x set V
T
is delivered every 10 to 100
breaths (sigh to breath ratio is
user selectable)