ESO 2 ICU Ventilator
25
User Manual
4. Pre-operation Procedures
4.1
Ventilator Set-up
WARNINGS
Before the ventilator is used clinically for the first time a
Calibration Check and Output Check must be successfully
completed. This work must be carried out by a Penlon-trained
technician, following the procedure in section 6.6 in the AV-S
service manual.
4.1.1 Electrical Power Connection
Before connecting the ventilator to the mains supply check that
the power supply is within the correct rating as stated on the
label on the rear of the control unit.
WARNING
Excessive electronic noise caused by other, poorly regulated
devices, such as electrocautery, may adversely interfere with
the proper functioning of the ventilator. Do not connect the
ventilator power cord into the same electrical wall outlet or
strip into which an electrocautery unit is connected.
Connection
Connect the mains cable to the socket (1) and plug into the
mains supply outlet.
WARNING
Check that the power cord can be easily disconnected.
Refer also to Section1, Warning 23.
Isolation from mains supply
Disconnect the cable from the mains supply outlet.
4.1.2 Ventilator Gas Supply
1. Check the drive gas supply (2) is air.
Always use Air.
2. Gas supply pressure range: 280 to 600 kPa (41 to 87 psi)
Supply pressure should be monitored by a separate means,
e.g. pressure gauge on the supply line.