ESO 2 ICU Ventilator
19
User Manual
Description
3.2.11 Spontaneous Mode and Patient Support Modes
CAUTION: No absorber in the rebreathing module!!
Spontaneous Mode
Default trigger level for ventilator support is 1.0 L/min.
Patient Support Modes
•
SIMV: Synchronised Intermittent Mandatory Ventilation
•
SMMV: Synchronised Mandatory Minute Ventilation
•
PSV: Pressure Supported Ventilation
For further information, see Appendix 5.
3.2.12 PEEP (Positive End Expiratory Pressure)
The AV-S ventilator includes a microprocessor-controlled, electronically integrated PEEP
system, regulated by the secondary pressure on the exhaust diaphragm (see Appendix 4).
The ventilator controls PEEP by allowing flow from, or delivering flow into, the bellows drive
circuit. This maintains the set pressure
NOTE
1. PEEP is electronically controlled.
2. PEEP is variable from 4 - 20 cmH2O (or optional 4 - 30 cmH2O), in increments of 1
cmH2O.
3. PEEP maximum limit is 20 cmH2O (or optional 30 cmH2O).
4. The display shows “OFF” when PEEP is not in use.
5. PEEP is switched off when the ventilator is switched off.
6. PEEP is switched off in Standby or ‘Spont’ mode to ensure minimal patient breathing
effort.
Selecting PEEP
1. Select by touching the screen tab PEEP, or using the navigator wheel. The setting will
flash.
2. Rotate the navigator wheel to set the required PEEP pressure.
A confirm message will be displayed.
Press the Screen Tab, or Wheel to confirm.
3. Note that
PEEP does not function in Spontaneous Mode or in any Patient Support Mode.
3.3
Spirometry
Enable or disable via the on-screen menu system.
If the spirometry system is turned OFF:
Fresh gas / fresh gas mixture compensation is disabled.
Support Modes are disabled.
Spirometry Menus
ON/OFF
1. Turn the navigator wheel to switch between ON and OFF.
2. Press to confirm.
3. Scroll to EXIT MENUS and press the wheel to exit.
Calibration
1. Press the navigator wheel to initiate the calibration procedure
(see section 4.1.7 for full procedure).
2. To exit the menu, scroll to EXIT MENUS and press the wheel.
Spirometry sub-menu - On/Off
O2 Monitor & Spiro
ESCAPE FROM MENU
O2 MONITOR: on
CALIBRATION: 100%
HIGH ALARM SET: 105
LOW ALARM SET: 18
>
SPIROMETER: on
SPIRO CALIBRATION: 0 L/min