ESO 2 ICU Ventilator
43
User Manual
7. Appendix
APPENDIX 1
Care of Back-up Battery
CAUTION
Damage may occur if the battery is allowed to remain in a
discharged state.
Battery installed in ventilator
The battery must be charged before the machine is released
for use with a 14-hour charge from the ventilator’s internal
power supply (ventilator connected to the mains supply, but
not running).
Note that the mains power indicator on the front panel will
show an amber light during charging.
Subsequently the recharge periods for a battery on a ventilator
in store are similar to those in B, below.
Batteries in machines in normal use will be kept charged by
the internal power supply.
Note that the Low Battery Alarm indicator may be displayed if
automatic recharging is taking place when the ventilator is in
use.
Battery care/storage requirements.
During storage, batteries will require a periodic recharge, the
frequency of which is determined by the storage temperature,
which must not exceed 50
o
C (120
o
F).
Storage temperature
Recharge period
38 to 50°C (100 to 122°F)
1 month
21 to 38°C (70 to 100°F)
3 months
7 to 21°C (45 to 70°F)
6 months
0 to 7°C (32 to 45°F)
9 months
-5 to 0°C (23 to 32°F)
12 months
Recharge duration - a charging cycle of at least 12 hours, to
ensure that the battery is kept at full capacity.
It is recommended that at each charge an updated label is
affixed to the unit to indicate date of the last charge.
Disposal of used batteries
Disposal of used batteries:
Do not dispose of in landfill, refer to an
approved recycling facility. Follow your hospital,
local, state and federal regulations.
NOTE
Removal/replacement of battery must only be undertaken by a
trained technician
APPENDIX 2
Disposal at end of useful life -
risk assessment
Do not dispose of this product in landfill, refer to
an approved recycling facility. Follow your hospital,
local, state and federal regulations.
EC territories: Follow the requirements of Directive 2002/96/
EC.
NOTE
Disposal of used batteries - see Appendix 1.
APPENDIX 3
Optional extras and approved
accessories
WARNING
Only use accessories approved by Penlon Ltd.
57655 Compact pressure tee
57523 Pressure sensing tube
57545 Adult bellows, base, and canister
57551 Adult canister
5006441 Adult bellows
57548 Bellows base
57553 Paediatric canister
57552 Paediatric bellows
57554 Paediatric bellows adaptor
UK Sales
Tel:
01235 547036
E-mail: [email protected]
International Sales
Tel:
+44 1235 547001
E-mail: [email protected]