ESO 2 ICU Ventilator
38
User Manual
Maintenance
6.2.7 Control Unit Patient Block Assembly
These operations must be carried out by suitably trained
technicians only.
6.2.7.1 Inspection: frequency / indications
On a regular basis (in line with hospital procedures for
infection control), and at least every six months, the patient
block (1) must be removed, cleaned and sterilised.
CAUTION:
After cleaning and sterilisation, and/or repair, the
patient block assembly must be refitted to the original control
unit (see operation 4 in 6.2.7.2, below).
6.2.7.2 Disassembly
1. Disconnect the drive gas and detach the hoses from the
outlets (2) on the outer face of the patient block assembly.
2. Note the different diameters for correct refitment.
3. Undo the securing knobs (3).
4. Carefully detach the assembly (1) from the control unit.
Note that resistance will be felt until the metal tubes (4)
and probe (5) disengage. Do not disassemble the patient
block.
NOTE
1. The serial number of the ventilator control unit (on the
label, 6) is also engraved on the patient block label (7).
2. The location of the ventilator control unit label (6) varies,
depending on the system layout.
3. After cleaning and sterilisation, the patient block must be
refitted to its original control unit.
6.2.7.3 Cleaning
1. Pre-cleaning: submerge the patient block in an enzymatic
solution within an ultrasonic tank for 20 minutes.
2. Clean the block in a washer/disinfector unit with an initial
cold rinse, a detergent wash, a decontamination stage at
92°C, followed by a final drying stage.
6.2.7.4 Sterilisation
1. Sterilise, as recommended in section 6.3. Do not
disassemble.
6.2.7.5 Reassembly
1. Important: The patient block must be refitted to the
original control unit.
The serial number on the patient block label (7) must
match the control unit serial number (on label, 6).
2. Position the patient block assembly (1) and push fully into
the control unit, ensuring that the metal tubes (4) are
engaged in their unions and the probe (5) locates in the
port in the patient block.
Fit the securing knobs (3).
6.2.7.6 Pre-use Checks
1. Function test the ventilator before clinical use (see 4.2.1.3)
.