Under
Advanced Settings
, select one of the three
Measurement Cycle
options, which are
Finite
, the acquisition will execute a number of cycles specified in # Cycles. The maximum
number of cycles is 9999,
Time Limit
, the acquisition will stop at a time and date set with the
Acquisition End
timestamp control
Ad Infinitum
, the acquisition will continue forever. Use the
STOP
button on the main user
interface to terminate the Lifetime experiment
Use the drop down menu of the
Acquisition Rate
ring control to set the rate of the measurement
cycle. The two available options are
1)
Maximum
SuperFACT will start a new acquisition as soon as the previous one is completed
2)
User
The User can force an idle time between two consecutive acquisitions. The interval is
set in
Meas.Interval
by entering it with the format hours:minutes:seconds
For lifetime run in conjunction with stress-biased experiments, the voltage bias to apply between
two consecutive acquisition cycles is set by selecting the
Bias On
option on the
Bias
drop down
menu, with the bias voltage specified in the
Bias Voltage
input field.
Bias Voltage
accepts voltage in
the range -10 to 10 Volts.
Note:
make sure that
Meas.Interval
is different from zero to allow the stress bias to be applied
Important:
If a total current measured during stress biased exceeds 10 mA, SuperFACT will return
error code 5600 and the acquisition will be terminated.
Tip:
Introduce an idle time between two consecutive acquisition cycles to reduce the amount of data
produced during a lifetime experiment