Lifetime and Stress Biased Acquisition
Stress biased acquisition
On the
Select Devices to Measure and Measurement Type
UI select
Stress Bias
and proceed to select the measurement type (TC and or IV) and the device to
measure etc. as for a standard acquisition. Select Apply to open the
Measurement Settings and Device Parameters
UI.
Under Advanced Settings, enter the number of measurement cycle in
# Cycles.
The minimum
number of cycles is 1, the maximum is 4
On the Acquisition Rate drop down menu, select User to force SuperFACT to wait a user defined
amount of time before starting the successive cycle.
Meas.Interval
is used to enter the interval between two successive acquisition cycles. This control
accepts value in the format hours:minutes:seconds.
The
Bias On
option of the
Bias
ring control drop down menu is used to enable the voltage bias
capability. The desired voltage bias is entered through the
Bias Voltage
input field.
Bias Voltage
accepts voltage in the range -10 to 10 Volts.
Important:
If the total current measured during stress biased exceeds 10 mA, SuperFACT will return
error code 5600 and the acquisition will be terminated.
Lifetime acquisition
On the
Select Devices to Measure and Measurement Type
UI select
Lifetime
to set a lifetime
experiment or
Lifetime
and
Stress Bias
to programme a lifetime combined with stress biased.
Proceed to select the measurement type (TC and or IV) and the device to measure etc. as for a
standard acquisition. Select
Apply
to open the
Measurement Settings and Device Parameters
UI.