Chapter 7 Maintenance and Repair
FLUSHING PUMP OFP-2
49
3.
Hold the front of the unit with one hand to prevent it from moving, then insert the fuse drawer
into the OFP-2 mains power inlet and push until it clicks into position.
CAUTION
Always ensure the fuse drawer is pushed fully home following
maintenance/repair, as failure to do so may cause the mains power
outlet to overheat.
4.
Plug in the power cord and turn the OFP-2 ON, confirm the power output.
WARNING
If the power fails to come on after replacing the fuses, unplug the power
immediately from the mains power inlet and contact Olympus.
CAUTION
If the fuses continue to blow, stop using the unit and contact Olympus or
hospital engineer.
7.5 Repair
The OFP-2 must only be serviced/repaired by suitably qualified technical personnel. If repair is
required, contact Olympus.
Olympus will not be responsible for damage or injury to equipment or personnel caused by
unauthorised repair or modification to this unit.
If the pump fails to operate:
Check the items detailed in “Troubleshooting” on page 54.
Check the condition of both fuses located in the fuse drawer in the mains power inlet
at the rear of the unit (see Section “Fuse Replacement” on page 48). Check the fuse
in the power cable plug as appropriate.
Figure 7.4