Chapter 3 Preparation for use
FLUSHING PUMP OFP-2
29
15.
Instrument channel connection (MAJ-1606/7)
: Refer to Figure 3.9 and connect the Luer
lock connector on the MAJ-1607 water tube to the MAJ
-
1606 instrument channel adaptor
attached to the instrument channel port on the endoscope.
16.
Auxiliary channel connection (with MAJ-1608/MAJ-855)
: Refer to Figure 3.10 and
connect the Luer lock connector on the MAJ-1608 water tube to the MAJ-855 auxiliary water
tube attached to the auxiliary channel port on the endoscope.
Figure 3.8
Figure 3.9
Pump head
lever
Pump head - side view
Tube
MAJ-1607
Endoscope
MAJ-1606
Instrument channel port
Instrument channel use
Содержание OFP-2
Страница 2: ......