Important Information — Please Read Before Use
FLUSHING PUMP OFP-2
11
• Inspect the MAJ-855 for obvious signs of wear or damage, and should
the slightest irregularity be suspected - do not use. Replace in
accordance with its instructions for use.
• Before each use, inspect the water container and cap for obvious signs
of wear or damage and should the slightest irregularity be suspected -
do not use.
• Always remove and empty the water container before moving the
equipment. Care should be taken to obtain a firm grip underneath the
body of equipment before moving. The water container must never be
used as a hand hold or carrying handle.
• Do not remove the water container while the pump is active. Ensure that
the pump is in the ‘OFF‘ position before removing or replacing the tube
sets.
• Store and use the OFP-2 within the environmental conditions described
in “Specification” on page 51; failure to do so may lead to equipment
malfunction or failure.
CAUTION
Only use replacement parts as specified in “Spares” on page 50. Failure
to do so may lead to equipment malfunction or failure.
• In accordance with the instructions supplied with the endoscope, ensure
that the auxiliary water channel or instrument channel is free from
blockages before starting a procedure.
• Personal protective equipment should be used in accordance with local
or national clinical guidelines.
• In the event of equipment failure, an alternative method of irrigation
should be used in accordance with the endoscope instruction manual.
• During use, do not obstruct the vents in the base and rear of the OFP-2
• The OFP-2 should not be positioned where the water container is
directly exposed to heat emitted from adjacent equipment, otherwise
fluid in the container may become hot after a period of time and scald
the patient.
• Before the procedure begins when using the MAJ-1681, ensure the
water container is filled with sterile saline.
• Activation of the pump while the tubing/EndoTherapy accessory is
connected to the OFP-2 but not inserted into an endoscope may result
in potentially contaminated fluid being pumped from the end of the
system and coming into contact with the floor, electrical equipment and
people in the area.
• If a water tube is installed incorrectly and becomes damaged, the water
tube set must be discarded and replaced.