Important Information — Please Read Before Use
FLUSHING PUMP OFP-2
10
• The Olympus OFP-2 is a medical device that requires safety measures
in regards to EMC. The device must be installed and put into operation
in accordance with the EMC information listed in “Operating
instructions” on page 33 of this document.
• Always use the supplied power cord. Use of any other power cord can
cause a malfunction of the device or the power cord to burn.
Additionally, the supplied power cord is for exclusive use with the device
it has been provided with and must not be used with other devices.
CAUTION
• Do not use sterile accessories if the packaging is damaged or opened.
• Use only in conjunction with OLYMPUS gastrointestinal/colono/
ultrasound endoscopes of EVIS EXERA/EXERA II/EXERA III/LUCERA/
LUCERA ELITE/EUS series and compatible EndoTherapy instruments.
CAUTION
• Check the “USE BY” date of all accessories as indicated on the
packaging label. Do not use any accessories beyond the specified “USE
BY” date.
• Do not allow the water container to run dry, otherwise air will be supplied
to the patient causing discomfort.
• To prevent free-flow of water to the patient, do not open the pump head
lever when the OFP-2 is connected to an endoscope which is inserted
into the patient. Ensure the OFP-2 is positioned below patient level.
• If any section of tube from the pump to the water container is blocked
or kinked, the flow rate will be noticeably reduced. Check the tube for
kinks or blockages, and should the slightest irregularity be suspected -
do not use.
• If a kinked or damaged tube causes the pump head to jam, the unit will
sense the increased load on the motor and shut down. Refer to
“Operating instructions” on page 33.
• Should the OFP-2 be operated with a blocked endoscope or MAJ-1608
with its tube pinch clamp closed, damage to the endoscope and/or water
tube may occur. In this event, care must be taken when disconnecting
the fluid path as the system may be pressurised. To assist in
depressurising the system, the pump head should be opened to allow
water back into the bottle. This must only be carried out with the
endoscope removed from the patient.
• In the event of tube pinch clamp failure when using MAJ-1608, drain the
fluid back to the source by opening the pump head and lifting the
connector end above the water level in the fluid container/ bottle before
disconnection. The tube and clamp should be replaced if any fault is
detected.
• Replace the tube and clean and sterilize the water container and water
container cap before the unit is reused.