Chapter 11 Technical description
FLUSHING PUMP OFP-2
59
11.5 Electromagnetic immunity
The OPF-2 is intended for use in the electromagnetic environment specified below. The customer
or the user of the OPF-2 should assure that it is used in such environment.
11.6 Accessories, transducers and cables
WARNING
Use of accessories, transducers and cables other than those specified
in this document could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in
improper operation.
Immunity
IEC 60601 test level
Electrostatic Discharge (ESD) IEC601000-4-
2:2008
± 8 kV contact, indirect
± 15 kV air
Electrical fast transient/burst IEC 61000-4-
4:2012
± 2 kV
for power supply lines
± 1 kV
for input/output lines
Surge IEC 61000-4-5:2014
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
Voltage dips, short interruptions and voltage
variations on power supply input lines IEC
61000-4-11:2004 + AMD1:2017
1) 0% UT , 1 cycle
2) 0% UT , ½ cycle,
Phase 0,45,90,135,180,225,270,315
3) 70% UT, (30% dip on UT for 25 cycles)
4) 0% UT for 5 sec
Conducted RF IEC 61000-4-6:2003 + A1:2004
+ A2:2006
6 Vrms ISM bands
3 Vrms
150 kHz to 80 MHz
Radiated RF IEC 61000-4-3:2006 +
AMD1:2007 + AMD2:2010
3 V/m
80 MHz to 2.7 GHz IEC 60601-1-2:Ed 4 Table 9
bands/levels/modulations.