Important Information — Please Read Before Use
FLUSHING PUMP OFP-2
9
• For EndoTherapy accessory channel flushing, use the MAJ-1681 or
MAJ-1682 accesssory port tubes, otherwise patient safety may be
compromised or damage to the pump may occur.
• When flushing with ø2.0 to ø2.2mm instrument channel videoscopes,
remove the EndoTherapy instrument from the MAJ-1606. Otherwise,
leakage or spraying of potentially contaminated fluid may occur. With an
EndoTherapy instrument inserted in the instrument channel, do not flush
and extract at the same time, as there is a potential risk of spraying.
WARNING
• If high flow rates are required while accessory tools are inserted in the
endoscope instrument channel, there is an increased risk of fluid
leakage. Therefore, the user should slow the pump speed to reduce any
fluid loss, or take precautionary measures to ensuring there is no
danger to users, patients or equipment.
• Note that the luer lock connector on the MAJ-855 includes a one-way
valve to prevent backflow - do NOT use the MAJ-855 without this
connector in place, or water tube, water container and water
contamination may result.
• Confirm that the wall power outlet has adequate electrical capacity to
power the equipment connected to it. Failure to do so may cause fire or
power failure to all equipment connected to the same power circuit.
• To avoid risk of electric shock, this equipment must only be connected to
a supply mains with protective earth.
• The water tube MAJ-1608 must be replaced daily - failure to do so may
lead to patient infection and a reduction in equipment performance.
• When removing the MAJ-855 from the MAJ-1608, firstly ensure the
pinch clamp is always closed to prevent fluid leakage when
disconnection is made, to avoid a slip hazard.
• Clean, disinfect or sterilize the MAJ-855 following each use according to
the manufacturer’s instructions supplied with the product and
instructions of the endoscope.
• The MAJ-1606, MAJ-1607, MAJ-1652, MAJ-1681 and MAJ-1682 are
single use, disposable items, and must be replaced after each patient.
Do not reuse or attempt to resterilize them. Failure to do so may lead to
risk of cross-infection.
• The Remote Control socket on the rear of the unit is for the exclusive
connection of compatible Olympus equipment, as shown in the "System
chart" in the Appendix. Connection to other equipment could
compromise the safety of the system.
• The operator should not touch the Remote Control socket whilst
simultaneously touching the patient.
• Do not modify this equipment.