Chapter 3 Preparation for use
FLUSHING PUMP OFP-2
22
CAUTION
• For endoscope, use only sterile water. For EndoTherapy device, use
only sterile saline. Sterile water should be degassed if the OFP-2 is to
be used with an ultrasound endoscope. Use of non-sterile fluid may
cause an infection risk to the patient.
• The water container should always be removed from the OFP-2 and
filled away from the workstation.
• To prevent spillage, take care not to tilt the container too far when filled
or when placing it on the water container tray. The water container will
leak fluid from the hole in the cap when tilted beyond 15°.
• To prevent fluid exuding from the cap:
- Take care not to squeeze the filled water container when
transferring back to the OFP-2.
- Take care when placing the filled water container into the
OFP-2 tray.
- Do not fill the water container past the “MAX” fill line.
6.
Insert the footswitch tube into the footswitch connection port on the front panel (Figure 3.2)
and place the footswitch on the floor.
7.
Route cabling neatly using the cable management provided on the workstation and connect
the IEC mains power cable to the mains power inlet on the rear of the unit and an easily
accessible, suitably grounded AC wall outlet or isolation/separation transformer socket.
Ensure that access to the mains inlet is not obstructed
WARNING
Do not allow the power cable plug to become wet, otherwise this may
cause electric shock.