Chapter 3 Preparation for use
FLUSHING PUMP OFP-2
23
NOTE
Refer to the workstation instructions for details on cable management
and use of the isolation/separation transformer.
8.
Referring to Figure 3.3, open the pump head lever on the pump head to access the pump
rollers.
9.
Confirm that the correct tube will be used for the procedure, noting the differences between
the tube sets.
WARNING
The MAJ-1607 instrument channel water tube, MAJ-1681 Accessory
Port Tube with Bottle Cap and MAJ-1682 Accessory Port Tube with
Saline Spike are single-use disposable items. The MAJ-1608 and
MAJ
-
1651 auxiliary channel water tube is a single day use item. Check
the model name shown on the sterile package and the tube, and use
them in accordance with their instruction manuals. Failure to do so may
pose an infection risk and/or cause equipment malfunction. Ensure
sterile tubing is free from damage. Do not use past the “use by“ date.
Figure 3.3
Model Name
Exchange frequency
Reason for exchange frequency
1. MAJ-1607
Instrument channel
water tube (2.5m)
After each patient use
The inside of the tube may be
contaminated when exchanging the
endoscope, it should be changed after
each patient use.
Disinfection and sterilization are not
available.
2. MAJ-1608
Auxiliary channel
water tube (0.75m)
After one day use
If the tube is not replaced after one day
use, it may cause contamination.
Disinfection and sterilization are not
available.
Pump head
Pump head lever