Chapter 3 Preparation for use
FLUSHING PUMP OFP-2
31
OFP-2 until all air bubbles have been expelled from the tubing.
NOTE
• In cold conditions, it may be necessary to prime the tubing at lower
pump speed.
• Priming is recommended between speed 5-9, depending on conditions.
• The MAJ-1681 and MAJ-1682 tubing must be primed with sterile saline
in order to fully expel the air already in the tubing. During this, the tubing
check valve may “whistle” until the air has been expelled. This is correct
operation and not a malfunction.
(iv) Refer to Figure 3.12, connect the Luer lock connector on the MAJ-1681 tubing to the
EndoTherapy device.
Set the pump speed to minimum.
WARNING
Use only in conjunction with compatible EndoTherapy devices listed in
the MAJ-1681/MAJ-1682 instruction manual.