Important Information — Please Read Before Use
FLUSHING PUMP OFP-2
8
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may
result in minor or moderate injury. It may also be used to alert against
unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.
Warnings and cautions
Follow the warnings and cautions given below when handling this equipment.
This information is to be supplemented by the warnings and cautions given in
each chapter. The OFP-2 is EMC-tested (electromagnetic compatibility) in
conformity with the requirements of IEC 60601-1-2:2014 Edition 4 and meets the
general electrical safety requirements of IEC 60601-1:2005 + A1:2012 Edition
3.1 (equivalent to EN 60601-1:2006 + A1:2013).
WARNING
• Explosion hazard - never install or use the OFP-2 within the zone of risk
of flammable gases.
• The OFP-2 should only be used in a medical facility under the direction
of a trained physician.
• For additional information, refer to the appropriate ‘instructions for use’
for the equipment and consumables used in conjunction with the OFP-2.
• The operator must assess the condition of the patient and use clinical
judgement to set the flow rate from the pump to a suitable level to avoid
patient trauma. The flow rate should always be checked at the start of
the procedure and be increased/decreased progressively to a level
commensurate with the clinical condition of the patient and degree of
washing required.
Note that this unit can deliver flow rates through the instrument
channel of the endoscope in excess of that which can be delivered
when using a 50ml syringe.
• The water container and cap were not cleaned, disinfected, or sterilized
before shipment. Before using the water container and the cap for the
first time, reprocess these according to “Operating instructions” on
• For endoscope instrument channel flushing, use only the MAJ-1607
instrument channel water tube with MAJ-1606 instrument channel
adaptor, otherwise patient safety may be compromised or damage to
the pump may occur.
• For endoscope auxiliary channel flushing, use the MAJ-1608 or MAJ-
1651/2 auxiliary channel water tube with MAJ
-
855 auxiliary water tube
to connect the OFP-2 to the endoscope, otherwise patient safety may
be compromised or damage to the pump may occur.